PerkinElmer Sprecturm2000 FT-IR (FT-IR-1)

From Help Wiki
Revision as of 15:59, 28 November 2012 by Geeraerd (Talk | contribs) (Created page with '<!-- FTIR-1 Instrument Page --> FT-IR-1.jpg <br/> <br/> [http://en.wikipedia.org/wiki/Fourier_Transform_InfraRed Fourier transform spectroscopy W…')

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

FT-IR-1.jpg

Fourier transform spectroscopy Wikipedia entry

Manual

Information Schema


Notes
  • DO NOT TURN OFF THE COMPUTER
  • DO NOT TURN OFF THE INSTRUMENT
  • If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on.
  • The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
  • Need to get software license key.
  • Software requires license key.
  • Software uses two directories found at the root of the drive: pel_apps & pel_data.
  • Software installation requires the 'mapping' or setting of the transputer I/O address. There's a help document at the root of the drive that explains how to do this.
  • The PC is a candidate to be upgraded, along with upgrading OS to Windows XP; it's likely that the software will not work on Windows 7, but worth testing in vm environment as a quick-and-dirty test.
  • The CD drive motor is broken.
  • Will not boot from USB.