Difference between revisions of "PerkinElmer Sprecturm2000 FT-IR (FT-IR-1)"
From Help Wiki
Line 3: | Line 3: | ||
<br/> | <br/> | ||
<br/> | <br/> | ||
− | [http://en.wikipedia.org/wiki/ | + | [http://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopy Fourier transform infrared spectroscopy Wikipedia entry] |
<br/> | <br/> | ||
<br/> | <br/> | ||
Line 13: | Line 13: | ||
<br/> | <br/> | ||
---- | ---- | ||
− | + | == Notes == | |
− | * DO NOT TURN OFF THE COMPUTER | + | ---- |
− | * DO NOT TURN OFF THE INSTRUMENT | + | * <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE COMPUTER!</span> |
− | * If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. | + | * <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE INSTRUMENT!</span> |
+ | * If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ''ready'' state. The ''Ready'' green LED light should be on after the instrument has initialized. | ||
+ | * The [PCI] computer controller board doens't automatically put the instrument into ''ready'' mode. Apparently, the [http://en.wikipedia.org/wiki/Industry_Standard_Architecture ISA] controller board does interact with the instrument to initialize it and put it into ''ready'' mode. | ||
* The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument. | * The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument. | ||
* Need to get software license key. | * Need to get software license key. | ||
Line 22: | Line 24: | ||
* Software uses two directories found at the root of the drive: pel_apps & pel_data. | * Software uses two directories found at the root of the drive: pel_apps & pel_data. | ||
* Software installation requires the 'mapping' or setting of the transputer I/O address. There's a help document at the root of the drive that explains how to do this. | * Software installation requires the 'mapping' or setting of the transputer I/O address. There's a help document at the root of the drive that explains how to do this. | ||
− | * The PC is a candidate to be upgraded, along with upgrading OS to Windows XP; it's likely that the software will not work on Windows 7, but worth testing in vm environment as a quick-and-dirty test. | + | * The instruments controller board has an EPROM with configuration; this configuration file has been saved and is on a floppy as well as being saved on the network share. |
+ | * The PC is a candidate to be upgraded, along with upgrading OS to Windows XP; it's likely that the software will not work on Windows 7, but worth testing in vm environment as a quick-and-dirty test. The OS needs to be 32 bit version. | ||
* The CD drive motor is broken. | * The CD drive motor is broken. | ||
* Will not boot from USB. | * Will not boot from USB. |
Revision as of 08:01, 14 May 2013
Fourier transform infrared spectroscopy Wikipedia entry
Manuals
Information Table
Notes
- DO NOT TURN OFF THE COMPUTER!
- DO NOT TURN OFF THE INSTRUMENT!
- If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ready state. The Ready green LED light should be on after the instrument has initialized.
- The [PCI] computer controller board doens't automatically put the instrument into ready mode. Apparently, the ISA controller board does interact with the instrument to initialize it and put it into ready mode.
- The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
- Need to get software license key.
- Software requires license key.
- Software uses two directories found at the root of the drive: pel_apps & pel_data.
- Software installation requires the 'mapping' or setting of the transputer I/O address. There's a help document at the root of the drive that explains how to do this.
- The instruments controller board has an EPROM with configuration; this configuration file has been saved and is on a floppy as well as being saved on the network share.
- The PC is a candidate to be upgraded, along with upgrading OS to Windows XP; it's likely that the software will not work on Windows 7, but worth testing in vm environment as a quick-and-dirty test. The OS needs to be 32 bit version.
- The CD drive motor is broken.
- Will not boot from USB.