Table List | Table Value |
---|---|
Wavelength Range | 190 to 820 nm |
Wavelength Accuracy | > ± 2 nm |
Wavelength Reproducibility | ± 0.05 nm (typical with constant conditions) |
Spectral Bandwidth | 2 nm |
Stray Light | at 200 nm = < 0.05% at 340 nm = < 0.05% |
Photometric Range | 0.0022 - 3.3 Å; 0.5 second measurements at 350 nm |
Photometric accuracy | 0.0022 - 3.3 Å; 0.5 second measurements at 350 nm |
Photometric Accuracy | ± 0.005 Å; at 1 Å, 440 nm, NIST 930e |
Photometric Noise | < 0.0002 Å RMS; at 0 Å, 500 nm, sixty 0.5-second scans |
Photometric Stability | < 0.001 Å/hr; (at 0 Å, constant ambient conditions, after 1 hour warm up): measured over 60 second period, every 5 seconds |
Baseline Flatness | < 0.001 Å RMS; 0.5 second blank followed by 0.5 second scan |
Shortest Scan Time | 0.1 seconds; full range |
Rapid Scan Mode | 10 scans per second; 70 nm range |
Power Requirements | Line Voltage: 80-132 or 170-264 V; AC Line Frequency: 47-66 Hz |
Dimensions | L: 23.8 in; W: 16.7 in; H: 7.3 in/td> |
Weight | 33.2 lbs. |
Environmental Conditions | Operating Temp.: 0-40 °C Non-operating Temp: -40 – 65 °C |