Difference between revisions of "PerkinElmer Sprecturm2000 FT-IR (FT-IR-1)"

From Help Wiki
(Created page with '<!-- FTIR-1 Instrument Page --> FT-IR-1.jpg <br/> <br/> [http://en.wikipedia.org/wiki/Fourier_Transform_InfraRed Fourier transform spectroscopy W…')
 
(update due to renaming)
 
(9 intermediate revisions by the same user not shown)
Line 1: Line 1:
<!-- FTIR-1 Instrument Page -->
+
<!-- PerkinElmer Sprecturm2000 FT-IR (FT-IR-1) Instrument Page -->
[[File:FT-IR-1.jpg|Right|Frame|FT-IR-1.jpg]]
+
[[File:FT-IR-1.jpg|FT-IR-1.jpg]]
 
<br/>
 
<br/>
 
<br/>
 
<br/>
[http://en.wikipedia.org/wiki/Fourier_Transform_InfraRed Fourier transform spectroscopy Wikipedia entry]
+
[http://en.wikipedia.org/wiki/Fourier_transform_infrared_spectroscopy Fourier transform infrared spectroscopy Wikipedia entry]
 
<br/>
 
<br/>
 
<br/>
 
<br/>
[[FTIR-1 Manual | Manual]]
+
[[FTIR-1 Manual | Manuals]]
 
<br/>
 
<br/>
 
<br/>
 
<br/>
[[FTIR-1 Information Schema | Information Schema]]
+
[[PerkinElmer Sprecturm2000 FT-IR (FT-IR-1) Information Table | Information Table]]
 
<br/>
 
<br/>
 
<br/>
 
<br/>
 
----
 
----
====== Notes ======
+
== Notes ==
* DO NOT TURN OFF THE COMPUTER
+
* <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE COMPUTER!</span>
* DO NOT TURN OFF THE INSTRUMENT
+
* <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE INSTRUMENT!</span>
* If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on.
+
* If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ''ready'' state. The ''Ready'' green LED light should be on after the instrument has initialized.
 +
* The [PCI] computer controller board doens't automatically put the instrument into ''ready'' mode. Apparently, the [http://en.wikipedia.org/wiki/Industry_Standard_Architecture ISA] controller board does interact with the instrument to initialize it and put it into ''ready'' mode. 
 
* The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
 
* The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
* Need to get software license key.
+
* Controller board information:
* Software requires license key.
+
** Perkin Elmer
* Software uses two directories found at the root of the drive: pel_apps & pel_data.
+
** PCI Single T4 PCB
* Software installation requires the 'mapping' or setting of the transputer I/O address. There's a help document at the root of the drive that explains how to do this.
+
** ASSY NO L1360664
* The PC is a candidate to be upgraded, along with upgrading OS to Windows XP; it's likely that the software will not work on Windows 7, but worth testing in vm environment as a quick-and-dirty test.
+
** ISS
* The CD drive motor is broken.
+
** "53-10/08/04"
* Will not boot from USB.
+
 
 +
* The instruments controller board has an EPROM with configuration: [[file:FTIR-Spectrum2000_eprom_config_41618.zip | EPROM configuration file]]
 +
----
 +
==== Computer & Software ====
 +
* Software requires license key which can be found on the back of the CD case or on the license certificate located in the CAL Manager file drawer.
 +
* Software uses two directories found at the root of the drive: pel_apps & pel_data. Standard user will need write access to "pel_data".
 +
* Software installation requires the 'mapping' or setting of the transputer I/O address; this is the PCI card. There's a help document on how to do this: [[File:Perkin_Elmer_Spectrum_Changing_the_Transputer_IO_Address.doc]]. An application will be required to set the memory address of the transputer: D:\IR Spectroscopy Tools\Set Transputer IO Address.exe
 +
* Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.
 +
 
 +
 
 +
===== Analysis Software =====
 +
* Use [http://www.effemm2.de/spekwin/index_en.html SpecWin32] software to analyze spectrum files outside of the Lab. This software is installed in the [http://www.evergreen.edu/scicomp/cal/hours-and-location.html CAL]
 +
 
 
<br/>
 
<br/>
 
<br/>
 
<br/>
 
----
 
----
[[Scientific Instrumentation | back to Main page]]
+
[[Scientific Instrumentation | back to '''Main''' page]]
 +
<br/>
 
[[Category:Scientific Instrumentation]]
 
[[Category:Scientific Instrumentation]]
 
<!-- END -->
 
<!-- END -->

Latest revision as of 09:54, 7 August 2017

FT-IR-1.jpg

Fourier transform infrared spectroscopy Wikipedia entry

Manuals

Information Table


Notes

  • DO NOT TURN OFF THE COMPUTER!
  • DO NOT TURN OFF THE INSTRUMENT!
  • If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ready state. The Ready green LED light should be on after the instrument has initialized.
  • The [PCI] computer controller board doens't automatically put the instrument into ready mode. Apparently, the ISA controller board does interact with the instrument to initialize it and put it into ready mode.
  • The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
  • Controller board information:
    • Perkin Elmer
    • PCI Single T4 PCB
    • ASSY NO L1360664
    • ISS
    • "53-10/08/04"

Computer & Software

  • Software requires license key which can be found on the back of the CD case or on the license certificate located in the CAL Manager file drawer.
  • Software uses two directories found at the root of the drive: pel_apps & pel_data. Standard user will need write access to "pel_data".
  • Software installation requires the 'mapping' or setting of the transputer I/O address; this is the PCI card. There's a help document on how to do this: File:Perkin Elmer Spectrum Changing the Transputer IO Address.doc. An application will be required to set the memory address of the transputer: D:\IR Spectroscopy Tools\Set Transputer IO Address.exe
  • Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.


Analysis Software
  • Use SpecWin32 software to analyze spectrum files outside of the Lab. This software is installed in the CAL




back to Main page