Difference between revisions of "PerkinElmer Sprecturm2000 FT-IR (FT-IR-1)"

From Help Wiki
(EPROM configuration file)
(update due to renaming)
 
(4 intermediate revisions by the same user not shown)
Line 1: Line 1:
<!-- FTIR-1 Instrument Page -->
+
<!-- PerkinElmer Sprecturm2000 FT-IR (FT-IR-1) Instrument Page -->
 
[[File:FT-IR-1.jpg|FT-IR-1.jpg]]
 
[[File:FT-IR-1.jpg|FT-IR-1.jpg]]
 
<br/>
 
<br/>
Line 9: Line 9:
 
<br/>
 
<br/>
 
<br/>
 
<br/>
[[FTIR-1 Information Table | Information Table]]
+
[[PerkinElmer Sprecturm2000 FT-IR (FT-IR-1) Information Table | Information Table]]
 
<br/>
 
<br/>
 
<br/>
 
<br/>
 
----
 
----
 
== Notes ==
 
== Notes ==
----
 
 
* <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE COMPUTER!</span>
 
* <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE COMPUTER!</span>
 
* <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE INSTRUMENT!</span>
 
* <span style="color: red; text-decoration: none; font-weight:bold; font-size: 14pt">DO NOT TURN OFF THE INSTRUMENT!</span>
Line 20: Line 19:
 
* The [PCI] computer controller board doens't automatically put the instrument into ''ready'' mode. Apparently, the [http://en.wikipedia.org/wiki/Industry_Standard_Architecture ISA] controller board does interact with the instrument to initialize it and put it into ''ready'' mode.   
 
* The [PCI] computer controller board doens't automatically put the instrument into ''ready'' mode. Apparently, the [http://en.wikipedia.org/wiki/Industry_Standard_Architecture ISA] controller board does interact with the instrument to initialize it and put it into ''ready'' mode.   
 
* The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
 
* The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
 +
* Controller board information:
 +
** Perkin Elmer
 +
** PCI Single T4 PCB
 +
** ASSY NO L1360664
 +
** ISS
 +
** "53-10/08/04"
 +
 
* The instruments controller board has an EPROM with configuration: [[file:FTIR-Spectrum2000_eprom_config_41618.zip | EPROM configuration file]]
 
* The instruments controller board has an EPROM with configuration: [[file:FTIR-Spectrum2000_eprom_config_41618.zip | EPROM configuration file]]
=== Computer & Software ===
+
----
* Need to get software license key. Software requires license key.
+
==== Computer & Software ====
* Software uses two directories found at the root of the drive: pel_apps & pel_data.
+
* Software requires license key which can be found on the back of the CD case or on the license certificate located in the CAL Manager file drawer.
* Software installation requires the 'mapping' or setting of the transputer I/O address. There's a help document at the root of the drive that explains how to do this.
+
* Software uses two directories found at the root of the drive: pel_apps & pel_data. Standard user will need write access to "pel_data".
* The PC is a candidate to be upgraded, along with upgrading OS to Windows XP; it's likely that the software will not work on Windows 7, but worth testing in vm environment as a quick-and-dirty test. The OS needs to be 32 bit version.
+
* Software installation requires the 'mapping' or setting of the transputer I/O address; this is the PCI card. There's a help document on how to do this: [[File:Perkin_Elmer_Spectrum_Changing_the_Transputer_IO_Address.doc]]. An application will be required to set the memory address of the transputer: D:\IR Spectroscopy Tools\Set Transputer IO Address.exe
* The CD drive motor is broken.
+
* Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.
* Will not boot from USB.
+
 
 +
 
 +
===== Analysis Software =====
 +
* Use [http://www.effemm2.de/spekwin/index_en.html SpecWin32] software to analyze spectrum files outside of the Lab. This software is installed in the [http://www.evergreen.edu/scicomp/cal/hours-and-location.html CAL]
 +
 
 
<br/>
 
<br/>
 
<br/>
 
<br/>

Latest revision as of 09:54, 7 August 2017

FT-IR-1.jpg

Fourier transform infrared spectroscopy Wikipedia entry

Manuals

Information Table


Notes

  • DO NOT TURN OFF THE COMPUTER!
  • DO NOT TURN OFF THE INSTRUMENT!
  • If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ready state. The Ready green LED light should be on after the instrument has initialized.
  • The [PCI] computer controller board doens't automatically put the instrument into ready mode. Apparently, the ISA controller board does interact with the instrument to initialize it and put it into ready mode.
  • The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
  • Controller board information:
    • Perkin Elmer
    • PCI Single T4 PCB
    • ASSY NO L1360664
    • ISS
    • "53-10/08/04"

Computer & Software

  • Software requires license key which can be found on the back of the CD case or on the license certificate located in the CAL Manager file drawer.
  • Software uses two directories found at the root of the drive: pel_apps & pel_data. Standard user will need write access to "pel_data".
  • Software installation requires the 'mapping' or setting of the transputer I/O address; this is the PCI card. There's a help document on how to do this: File:Perkin Elmer Spectrum Changing the Transputer IO Address.doc. An application will be required to set the memory address of the transputer: D:\IR Spectroscopy Tools\Set Transputer IO Address.exe
  • Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.


Analysis Software
  • Use SpecWin32 software to analyze spectrum files outside of the Lab. This software is installed in the CAL




back to Main page