* Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.
 
* Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.
  
 +
 +
===== Analysis Software =====
 +
* Use [http://www.effemm2.de/spekwin/index_en.html SpecWin32] software to analyze spectrum files outside of the Lab. This software is installed in the [http://www.evergreen.edu/scicomp/cal/hours-and-location.html CAL]
  
 
<br/>
 
<br/>

Revision as of 09:37, 31 October 2013

FT-IR-1.jpg

Fourier transform infrared spectroscopy Wikipedia entry

Manuals

Information Table


Notes

  • DO NOT TURN OFF THE COMPUTER!
  • DO NOT TURN OFF THE INSTRUMENT!
  • If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ready state. The Ready green LED light should be on after the instrument has initialized.
  • The [PCI] computer controller board doens't automatically put the instrument into ready mode. Apparently, the ISA controller board does interact with the instrument to initialize it and put it into ready mode.
  • The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
  • The instruments controller board has an EPROM with configuration: File:FTIR-Spectrum2000 eprom config 41618.zip

Computer & Software

  • Software requires license key which can be found on the back of the CD case or on the license certificate located in the CAL Manager file drawer.
  • Software uses two directories found at the root of the drive: pel_apps & pel_data. Standard user will need write access to "pel_data".
  • Software installation requires the 'mapping' or setting of the transputer I/O address; this is the PCI card. There's a help document on how to do this: File:Perkin Elmer Spectrum Changing the Transputer IO Address.doc. An application will be required to set the memory address of the transputer: D:\IR Spectroscopy Tools\Set Transputer IO Address.exe
  • Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.


Analysis Software
  • Use SpecWin32 software to analyze spectrum files outside of the Lab. This software is installed in the CAL