					| Table List | Table Value | 
|---|---|
| Wavelength Range | 190 to 820 nm | 
| Wavelength Accuracy | > ± 2 nm | 
| Wavelength Reproducibility | ± 0.05 nm (typical with constant conditions) | 
| Spectral Bandwidth | 2 nm | 
| Stray Light | at 200 nm = < 0.05% at 340 nm = < 0.05% | 
| Photometric Range | 0.0022 - 3.3 Å; 0.5 second measurements at 350 nm | 
| Photometric accuracy | 0.0022 - 3.3 Å; 0.5 second measurements at 350 nm | 
| Photometric Accuracy | ± 0.005 Å; at 1 Å, 440 nm, NIST 930e | 
| Photometric Noise | < 0.0002 Å RMS; at 0 Å, 500 nm, sixty 0.5-second scans | 
| Photometric Stability | < 0.001 Å/hr; (at 0 Å, constant ambient conditions, after 1 hour warm up): measured over 60 second period, every 5 seconds | 
| Baseline Flatness | < 0.001 Å RMS; 0.5 second blank followed by 0.5 second scan | 
| Shortest Scan Time | 0.1 seconds; full range | 
| Rapid Scan Mode | 10 scans per second; 70 nm range | 
| Power Requirements | Line Voltage: 80-132 or 170-264 V; AC Line Frequency: 47-66 Hz | 
| Dimensions | L: 23.8 in; W: 16.7 in; H: 7.3 in/td> | 
| Weight | 33.2 lbs. | 
| Environmental Conditions | Operating Temp.: 0-40 °C Non-operating Temp: -40 – 65 °C | 