Difference between revisions of "PerkinElmer Sprecturm2000 FT-IR (FT-IR-1)"

From Help Wiki
m (Geeraerd moved page FTIR-1 to PerkinElmer Sprecturm2000 FT-IR (FT-IR-1): Following naming convention)
(update due to renaming)
 
Line 1: Line 1:
<!-- FTIR-1 Instrument Page -->
+
<!-- PerkinElmer Sprecturm2000 FT-IR (FT-IR-1) Instrument Page -->
 
[[File:FT-IR-1.jpg|FT-IR-1.jpg]]
 
[[File:FT-IR-1.jpg|FT-IR-1.jpg]]
 
<br/>
 
<br/>
Line 9: Line 9:
 
<br/>
 
<br/>
 
<br/>
 
<br/>
[[FTIR-1 Information Table | Information Table]]
+
[[PerkinElmer Sprecturm2000 FT-IR (FT-IR-1) Information Table | Information Table]]
 
<br/>
 
<br/>
 
<br/>
 
<br/>
Line 24: Line 24:
 
** ASSY NO L1360664
 
** ASSY NO L1360664
 
** ISS
 
** ISS
** 53-10/08/04
+
** "53-10/08/04"
  
 
* The instruments controller board has an EPROM with configuration: [[file:FTIR-Spectrum2000_eprom_config_41618.zip | EPROM configuration file]]
 
* The instruments controller board has an EPROM with configuration: [[file:FTIR-Spectrum2000_eprom_config_41618.zip | EPROM configuration file]]

Latest revision as of 09:54, 7 August 2017

FT-IR-1.jpg

Fourier transform infrared spectroscopy Wikipedia entry

Manuals

Information Table


Notes

  • DO NOT TURN OFF THE COMPUTER!
  • DO NOT TURN OFF THE INSTRUMENT!
  • If the computer is turned off, the spectrum software will need to be run; do a 'BkGrd' (background scan) from the home screen. This needs to be done every time the computer is turned back on. This will initialize the instrument and put it into a ready state. The Ready green LED light should be on after the instrument has initialized.
  • The [PCI] computer controller board doens't automatically put the instrument into ready mode. Apparently, the ISA controller board does interact with the instrument to initialize it and put it into ready mode.
  • The instrument should be left on so the internal heater can keep the inside of the instrument dry. Internal moisture will degrade the salt based piece inside the instrument.
  • Controller board information:
    • Perkin Elmer
    • PCI Single T4 PCB
    • ASSY NO L1360664
    • ISS
    • "53-10/08/04"

Computer & Software

  • Software requires license key which can be found on the back of the CD case or on the license certificate located in the CAL Manager file drawer.
  • Software uses two directories found at the root of the drive: pel_apps & pel_data. Standard user will need write access to "pel_data".
  • Software installation requires the 'mapping' or setting of the transputer I/O address; this is the PCI card. There's a help document on how to do this: File:Perkin Elmer Spectrum Changing the Transputer IO Address.doc. An application will be required to set the memory address of the transputer: D:\IR Spectroscopy Tools\Set Transputer IO Address.exe
  • Perkin Elmer Spectroscopy software version 5.3.1 will not work on Windows 7 --already tested, and also not supported. Also, the software will only work on Windows XP 32 bit edition.


Analysis Software
  • Use SpecWin32 software to analyze spectrum files outside of the Lab. This software is installed in the CAL




back to Main page